Technical Parameters Of Blood Oxygen Pulse Rate
Blood oxygen saturation
Blood oxygen saturation measurement range: 0%-100% Blood oxygen saturation display resolution: 1 %
The accuracy of blood oxygen saturation measurement: in the range of 70% -100%, the measurement error should be ±4%; in the range of 0% – 69%, the measurement error is not defined.
Blood oxygen saturation alarm setting range: the upper limit of the alarm should be 1%-100%; the lower limit of the alarm should be 0%-99%
Blood oxygen saturation alarm accuracy: the alarm accuracy should not exceed ± 1 % of the alarm setting value. Technical alarm: there should be an alarm when the sensor falls off
Pulse rate: The pulse rate detection range should be 30bpm-240bpm
Pulse rate detection resolution: 1 bpm;
The accuracy of pulse rate measurement should be ±3bpm;
Pulse rate alarm setting range: the upper limit of the alarm should be 2bpm-254bpm;
The lower alarm limit should be 0bpm-252bpm
Pulse rate alarm accuracy: the alarm accuracy should not exceed the set value ± 1 bpm.
Built-In Side Stream C02 Technical Parameters
Sensor type: side stream COZ
Sampling flow: SO -1 S0ml/min; default 75ml/min
Technical principie: dual-wavelength non-dispersive infrared (NDIR), no moving parts initialization time: display the waveform within 3 seconds, reach the index requirement within 5 minutes (at 2SºC room temperature)
Carbon dioxide range: O – 114 mmHg, O – 15 %, O – 15.2 kPa Rise time: less than 4s
Carbon dioxide resolution:
• mmHg O – 38 mmHg
• 0.25 mmHg 39 – 114mmHg
Carbon dioxide accuracy: O – 40mmHg ±2 mmHg
• 41 – 76 mmHg ±5% of reading
• 77 – 114 mmHg ±8% of reading
Respiration rate greater than 80 BPM ±12% of reading Carbon dioxide stability: short-term drift: less than 1 mmHg within 4 hours
Long-term drift: meet the measurement accuracy requirements within 120 hours
Data sampling rate: 100 Hz
Respiration rate range: 3-120 times/min
Respiration rate accuracy: ±1 times/minute
Compensation: Atmospheric pressure range: 400 mmHg-800 mmHg (automatic compensation)
Calibration: The user does not need to calibrate regularly. Voltage requirements: 5.0 VDC ±5%, 2W
Power consumption: 1.0W (typical value); 1.8W maximum (warm-up)
Data interface: RS232, 19200 baud rate, Standard N-8-1
Data output: carbon dioxide concentration curve, end-tidal carbon dioxide concentration, inhaled carbon dioxide concentration, respiration rate. (The host is equipped with a mixed gas type, and the measurement result is the compensated value)
Non-Invasive Blood Pressure Technical Parameters
Static pressure measurement range: Static pressure measurement range 0kPa (0mmHg) – 40kPa (300mmHg), measurement accuracy ±3mmHg (±0.4kPa)
The accuracy of systolic blood pressure, mean blood pressure and diastolic blood pressure:
• Maximum average error ±0.67kPa (±SmmHg)
• Maximum standard deviation 1.067kPa (8mmHg)
Range: The range is at least 0kPa (0mmHg) to 34.67kPa (260mmHg)
Resolution: Display resolution: 0.133kPa (1 mmHg)
Repeatability: Measured in the static continuous low pressure state, the difference between the repeated measurement readings at each point within the scale range should not be more than 0.533kPa (4mmHg).
Pressure sensor accuracy: Regardless of whether the pressure is increased or decreased, the maximum error of the pressure measurement in the cuff should be ±0.4kPa (±3mmHg) at any test point in the range
Technical Parameters Of Blood Oxygen Pulse Rate
Blood oxygen saturation
Blood oxygen saturation measurement range: 0%-100% Blood oxygen saturation display resolution: 1 %
The accuracy of blood oxygen saturation measurement: in the range of 70% -100%, the measurement error should be ±4%; in the range of 0% – 69%, the measurement error is not defined.
Blood oxygen saturation alarm setting range: the upper limit of the alarm should be 1%-100%; the lower limit of the alarm should be 0%-99%
Blood oxygen saturation alarm accuracy: the alarm accuracy should not exceed ± 1 % of the alarm setting value. Technical alarm: there should be an alarm when the sensor falls off
Pulse rate: The pulse rate detection range should be 30bpm-240bpm
Pulse rate detection resolution: 1 bpm;
The accuracy of pulse rate measurement should be ±3bpm;
Pulse rate alarm setting range: the upper limit of the alarm should be 2bpm-254bpm;
The lower alarm limit should be 0bpm-252bpm
Pulse rate alarm accuracy: the alarm accuracy should not exceed the set value ± 1 bpm.