1. Monoblock
1.1 Focusing: 0.6 / 1.8
1.2 Anode Capacity: 35.5 kJ (47 kHu)
1.3 Tube Heat Capacity: 650 kJ (867 kHu)
1.4 Output Power: 5 kW
1.5 Inverter Frequency: ≥ 40 kHz
1.6 Continuous Fluoroscopy (Manual/Automatic)
1.6.1 Tube Voltage: 40 kV – 120 kV
1.6.2 Tube Current: 0.3 mA – 4 mA
1.6.3 Automatic Brightness Tracking Function
1.7 Pulsed Fluoroscopy
1.7.1 Tube Voltage: 40 kV – 120 kV
1.7.2 Tube Current: 0.3 mA – 30 mA
1.8 X-ray Mode
1.8.1 X-ray: 40 kV – 120 kV
1.8.2 X-ray tube current: 25 mA – 100 mA
1.8.3 X-ray mAs: 1.0 mAs – 180 mAs
1.9 Beam limiter: Electric iris + symmetrical linear rotary
1.10 Working environment conditions
1.10.1 Ambient temperature: 10 °C – 40 °C
1.10.2 Relative humidity: 30 % – 75 %
1.10.3 Atmospheric pressure: 700 hPa – 1060 hPa
1.11 Power operating conditions
1.11.1 Supply voltage and number of phases: single-phase 220 V ± 22 V
1.11.2 Supply frequency: 50 Hz ± 1 Hz
1.11.3 Internal resistance of the power supply: no more 1 Ω
2. Imaging System
2.1 Image Intensifier: 9″, three fields e5764sd-p3, center resolution 4.8 lp/mm
2.2 Ultra-low illuminance megapixel progressive scan black and white camera, pixel array description: 1024 x 1024
2.3 LCD: 24-inch standard LCD, resolution 1920 x 1200, operating frequency: 60 Hz
2.4 Image Acquisition and Processing Workstation
2.4.1 Registration: Record retention, medical history query, worklist
2.4.2 Acquisition: Start acquisition, prepare recording, restart, horizontal mirror, vertical mirror, window adjustment, magnifying glass, negative image, edge enhancement, recursive noise reduction
2.4.3 Processing: Four windows, nine windows, sharpening, horizontal mirror, vertical mirror, text annotation, image measurement Length
2.4.4 Report: Save, Preview, Expert Template
2.4.5 DICOM Function: DICOM Scan, Network Service
2.5 Image Definition Index
2.5.1 Gray Level: ≥12
3. Mechanical Components
3.1 Forward and Backward Movement: 200 mm
3.2 Rotation Around the Horizontal Axis: ±180°
3.3 Rotation Around the Vertical Axis: ±15°
3.4 Focal Length: 960 mm
3.5 C-Arm Opening: 760 mm
3.6 C-Arm Arc Depth: 640 mm
3.7 Sliding Along the Track: 120° (+90° ~ -30°)
3.8 Electric Column Lift: 400 mm
3.9 Guide Wheel and Main Wheel: The guide wheel can rotate in any direction, and the main wheel Main monitor can rotate ±90°
3.10 Monitor rotation on frame: ≥300°
3.11 Luminous push
Features
1. A high-quality, high-frequency, high-voltage combined X-ray generator significantly reduces X-ray exposure.
2. It features automatic kV and mA tracking, which automatically adjusts image brightness and clarity to maintain optimal image quality.
3. A graphic LCD touch screen main operating interface is adopted for smarter and more convenient operation.
4. The handheld controller design makes the instrument easy to operate.
5. A 9-inch, three-field image intensifier is used, with stable and reliable image quality and good image definition.
6. An ultra-low-light, megapixel digital camera is used, providing a clearer image.
7. The standard workstation and advanced image processing technology make the image clearer, facilitating operation and diagnosis by doctors.
It features a standard DICOM interface and easy integration with the hospital information system.
8. New frame design, small and aesthetically pleasing.
9. Supports digital photography, making photography operations more convenient and digital image processing more efficient.