Sterilizers Horizontal – STERI ARIES STEAM 372L

  • Categoria Medica

Detalles

  • Marca Welch Medical
  • Modelo Steri Aries Steam 372L
  • Fabricante USA
  • Origen USA

Especificaciones Tecnicas

– Volume: 300L and 372L (Internal chamber dimensions for the 372L model are 610 x 610 x 1000 mm).
– Structure design: 316 stainless steel with a service life of 15 years or 30,000 cycles.
– Operating temperature: 121°C to 139°C (with a maximum design temperature of 150°C).
– Maximum working pressure: 0.25 MPa.
– Design pressures: -0.1/0.3 MPa for the internal chamber and 0.3 MPa for the jacket.
– Parameter accuracy: Sterilization temperature control of ±0-2°C, temperature uniformity of ±1°C, display accuracy of 0.5°C for temperature and 0.1 kPa for pressure.
– Vacuum System: Maximum vacuum of -96 kPa and a leakage rate less than or equal to 0.13 kPa/min.
– Pulse Cycles: 3 times by default (adjustable from 0 to 400 times).
– Supported Programs: Features 13 cycles including sterile fabrics/dressings, conventional instruments, a quick (flash) program, uncovered liquids, Bowie & Dick test, vacuum leak test, PCD test, and drying.
– Electrical Requirements: AC 380V ± 38V, 50Hz ± 1Hz.
– Equipment Power: 2 kW+ (30 kW with steam generator).
– Steam Consumption: 45 kg/°C.
– User Interface: 7-inch or larger touchscreen, with built-in thermal printer, data logger, and multi-language options (English, Spanish, etc.).

Descripcion

Reinforced Structure: Features a European-style double-layer annular jacket design that reduces stress concentration and dramatically increases load capacity and equipment lifespan under fluctuating temperature and pressure loads.
High-End Components: Equipped with a Siemens control system (PLC and touchscreen), German temperature probe, pressure transmitter, angle seat valve, and a SPECK liquid ring vacuum pump.
Advanced Door Safety: Equipped with a motorized door system with safety interlocks based on both pressure (prevents opening if the pressure exceeds 20 kPa) and temperature (in liquid programs, the door will not open until the temperature drops below 60°C).
Use and Application: Ideal for sterilizing moisture- and heat-resistant medical materials, including solid and hollow surgical instruments, fabric dressings, rubber items, and unsealed liquids.

rrss