– Tidal Volume: Range of 20 to 1500 mL, suitable for neonatal, pediatric, and adult patients.
– Respiratory Rate: Adjustable from 4 to 100 breaths per minute for precise adaptive control.
– I:E Ratio: Flexible configuration with ratios from 4:1 to 1:8.
– Maximum Pressure (Pmax): Adjustable safety limit between 10 and 80 cmH2O to prevent barotrauma.
– Electronic PEEP: Integrated control with an operating range of 0 to 20 cmH2O.
– Gas Inlet: System for O2, N2O, and air with standard pressures of 2.8 to 6 bar.
– Hypoxic Guard: Mechanical/electronic protection that guarantees a minimum of 25% O2 in the mixture.
– Vaporizer Support: System compatible with dual Selectatec and a locking mechanism. Autonomy: Integrated backup battery with a runtime of 90 to 120 minutes.
– Connectivity: RS232 and Ethernet ports for data export and external monitoring.
• Integrated Anesthesia Workstation: Designed for adult, pediatric, and neonatal patients with high precision.
• Electronic Ventilator: Incorporates an electronically controlled ventilator powered by pneumatic components.
• Versatile Ventilation Modes: Offers advanced modes such as VCV, PCV, SIMV, PSV, and manual/spontaneous ventilation.
• Color TFT Display: High-resolution display (typically 10.4 to 12.1 inches) for real-time monitoring.
• Real-Time Graphs: Simultaneously displays pressure, flow, and volume curves for optimal clinical control.
• Electronic/Mechanical Flowmeters: High-precision system for the accurate delivery of O₂, N₂O, and air.
• Integrated Gas Monitoring: Capable of measuring CO₂ and anesthetic gases (halothane, isoflurane, sevoflurane, etc.).
• Compact Suction System: Integrated breathing circuit that is autoclavable and easy to disassemble to prevent infection. • Volume compensation: Automatic system that corrects tidal volume according to circuit compliance.
• Integrated safety (Hypoxic Guard): Protection mechanism that guarantees a minimum oxygen concentration of 25%.
• Backup battery: Long-lasting internal battery to ensure continuous operation in the event of power outages.
• Modular and ergonomic design: Robust structure with storage drawers and rails for additional monitors.
• Selectatec compatibility: Dual vaporizer support with an interlocking system for enhanced safety.
• Intelligent alarms: Configurable system of audible and visual alerts for physiological and equipment parameters.
• Low maintenance costs: Highly durable components that optimize operating costs in the hospital environment.